Model 14 Oxygen Analyzer
7-4
Interferences Are there any sources of potential
interference, chemical, external,
temperature, pressure, etc.?
Cable shield is not connected Check and connect
Defective CPU board If problem persists, call the Service Technician
Temperature measurement Probe connected wrong Check probe connection
is not correct
Temperature not calibrated Calibrate for temperature. Also check the
T.C. for correct resistance (ohmmeter)
CPU board is defective If problem persists, call the Service Technician
Display reading static; CPU board is malfunctioning and/or Check probe connection
cannot be changed in some other transmitter hardware is
any way defective Initiate a software RESET
If Steps 1-3 do not remedy the problem,
make a cold start (RESET): interrupt the
power for 5-10 seconds
Reprogram the instrument
If problem persists, call the Service Technician
Relays not energized Instrument was programmed Check whether the correct relay parameters
incorrectly and set points have been programmed
Hardware is defective Check that the programmed set points are
compatible with the programmed measuring
range
Check the relay characteristics for proper
functioning using an ohmmeter
If problem persists, contact the Service
Technician
Wrong output current, Instrument was incorrectly Check the programmed output-current
output current remains programmed parameters
locked at 0 or 20 mA
Connection of the Model 14 with Check the cables
peripherals (recorder, etc.) are faulty,
loose or defective
Hardware is defective Compare the measured value with the
output-current range
If problem persists, call the Service Technician
Polarization voltage Wrong configuration (3 electrode Check that the switches under the
incorrect mode) amperometric board are correctly positioned
(“on”)
Wrong programming Check programmed parameters
MALFUNCTION POSSIBLE CAUSE REMEDIES