Model 14 Oxygen Analyzer
5-6
SYSTEM ALARM FUNCTION:
Relays S3 and S4 may be used as a fault indicater. To control the faults traced by the analyzer, connect the
specific relay to an external alarm system. The relay is activated as soon as a default appears.
In the case of a manual acknowledgment, the relay remains activated even if the default disappears.
Press ENTER to deactivate the relay and the error message. In the case of an automatic acknowledgment,
the relay and the error message are deactivated when the default disappears.
ALARM 3
Select
MODE: System
ACCEPT.: Auto
RELAY: NC
MODE -No The alarm S3 may be programmed to be a limit
-Limit function (see paragraph above) or an alarm
-System system function
ACCEPT -Auto In the case of an alarm system, choose between
-Manu. a manual (key ENTER) or an automatic
acknowledgment
RELAY -NO Alarm S3 can be normally open or normally closed
-NC
System Alarm